User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Kistler 5010:Charge Amplifier DATE: 17-Dec-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 31 NUMBER OF LINES: 287 CONFIGURATION: Fluke 5700A CONFIGURATION: Datron 1281 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P U M F W 1.002 ASK+ X 1.003 IEEE [@1281]*RST 1.004 HEAD INITIAL CONDITIONS 1.005 DISP Connect the UUT CHARGE input BNC through the 1000pF 1.005 DISP standard capacitor to the 5700 V ê A outputs. 1.006 DISP Connect the UUT OUTPUT connector to the 1281 1.006 DISP HI/LO inputs. 1.007 HEAD NOTE !!!! 1.008 DISP As you use the SELECT and arrows during this procedure 1.008 DISP the UUT will display RESET? YES. At this point press the 1.008 DISP SELECT button and continue using the arrow keys as 1.008 DISP necessary. 1.009 HEAD {* Scale Accuracy Tests *} 1.010 JMP 2.001 1.011 EVAL 2.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 2.001 DISP TRANSDUCER SENSITIVITY for 9.99E+3 pc/MU. 2.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 2.002 DISP SCALE .01 MU/Volt. 2.003 DISP Set the UUT MODE to CHARGE, TIME CONSTANT to LONG, 2.003 DISP and to OPERATE. 2.004 IEEE [@1281]ACV AUTO 2.005 5700 7.07mV 400H O S 2W 2.006 IEEE [@1281]RDG?[D2000][I] 2.007 MEM* 1000 2.008 ACC 70.7mV 0.03% 2.009 MEME 2.010 MEMC mV 0.50000% .01MU/Volt #! Test Tol 0.3535, Sys Tol 0.02121, TUR 16.667 (>= 4.00). 3.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 3.001 DISP TRANSDUCER SENSITIVITY for 1.00E+3 pc/MU. 3.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 3.002 DISP SCALE .1 MU/Volt. 3.003 DISP Set the UUT to OPERATE. 3.004 5700 70.7mV 400H S 2W 3.005 IEEE [@1281]RDG?[D2000][I] 3.006 ACC .707V 0.03% 3.007 MEME 3.008 MEMC V 0.50000% .1MU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 4.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 4.001 DISP TRANSDUCER SENSITIVITY for 1.00E+2 pc/MU. 4.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 4.002 DISP SCALE 1 MU/Volt. 4.003 DISP Set the UUT to OPERATE. 4.004 5700 70.7mV 400H S 2W 4.005 IEEE [@1281]RDG?[D2000][I] 4.006 ACC .707V 0.03% 4.007 MEME 4.008 MEMC V 0.50000% 1MU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 5.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 5.001 DISP SCALE 10 MU/Volt. 5.002 DISP Set the UUT to OPERATE. 5.003 5700 707mV 400H S 2W 5.004 IEEE [@1281]RDG?[D2000][I] 5.005 ACC .707V 0.03% 5.006 MEME 5.007 MEMC V 0.50000% 10MU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 6.001 HEAD Scale Accuracy Tests 6.002 JMP 7.001 6.003 EVAL 7.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 7.001 DISP TRANSDUCER SENSITIVITY for 1.00E+1 pc/MU. 7.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 7.002 DISP SCALE 100 MU/Volt. 7.003 DISP Set the UUT to OPERATE. 7.004 5700 707.0mV 400H S 2W 7.005 IEEE [@1281]RDG?[D2000][I] 7.006 ACC .707V 0.03% 7.007 MEME 7.008 MEMC V 0.50000% 100MU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 8.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 8.001 DISP SCALE 200 MU/Volt. 8.002 DISP Set the UUT to OPERATE. 8.003 5700 1.414V 400H S 2W 8.004 IEEE [@1281]RDG?[D2000][I] 8.005 ACC .707V 0.03% 8.006 MEME 8.007 MEMC V 0.50000% 200MU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 9.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 9.001 DISP SCALE 500 MU/Volt. 9.002 5700 3.535V 400H S 2W 9.003 IEEE [@1281]RDG?[D2000][I] 9.004 ACC .707V 0.03% 9.005 MEME 9.006 MEMC V 0.50000% 500MU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 10.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 10.001 DISP SCALE 1000 MU/Volt. 10.002 5700 7.07V 400H S 2W 10.003 IEEE [@1281]RDG?[D2000][I] 10.004 ACC .707V 0.03% 10.005 MEME 10.006 MEMC V 0.50000% 1kÿMU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 11.001 HEAD Scale Accuracy Tests 11.002 JMP 12.001 11.003 EVAL 12.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 12.001 DISP TRANSDUCER SENSITIVITY for 1.00E+0 pc/MU. 12.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 12.002 DISP SCALE 10000 MU/Volt. 12.003 DISP Set the UUT OPERATE. 12.004 5700 7.07V 400H S 2W 12.005 IEEE [@1281]RDG?[D2000][I] 12.006 ACC .707V 0.03% 12.007 MEME 12.008 MEMC V 0.50000% 10kÿMU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 13.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 13.001 DISP TRANSDUCER SENSITIVITY for 1.00E-1 pc/MU. 13.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 13.002 DISP SCALE 100000 MU/Volt. 13.003 DISP Set the UUT OPERATE. 13.004 5700 7.07V 400H S 2W 13.005 IEEE [@1281]RDG?[D2000][I] 13.006 ACC .707V 0.03% 13.007 MEME 13.008 MEMC V 0.50000% 100kÿMU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 14.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 14.001 DISP TRANSDUCER SENSITIVITY for 1.00E-2 pc/MU. 14.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 14.002 DISP SCALE 1M MU/Volt. 14.003 DISP Set the UUT OPERATE. 14.004 5700 7.07V 400H S 2W 14.005 IEEE [@1281]RDG?[D2000][I] 14.006 ACC .707V 0.03% 14.007 MEME 14.008 MEMC V 0.50000% 1MÿMU/Volt #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 15.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 15.001 DISP SCALE 10M MU/Volt. 15.002 DISP Set the UUT OPERATE. 15.003 5700 7.07V 400H S 2W 15.004 IEEE [@1281]RDG?[D2000][I] 15.005 ACC .0707V 0.03% 15.006 MEME 15.007 MEMC V 0.50000% 10MÿMU/Volt #! Test Tol 0.0003535, Sys Tol 2.121e-005, TUR 16.667 (>= 4.00). 16.001 HEAD {* Transducer Sensitivity Tests *} 16.002 JMP 17.001 16.003 EVAL 17.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 17.001 DISP TRANSDUCER SENSITIVITY for 1.00E+3 pc/MU. 17.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 17.002 DISP SCALE .1 MU/Volt. 17.003 DISP Set the UUT OPERATE. 17.004 5700 70.7mV 400H S 2W 17.005 IEEE [@1281]RDG?[D2000][I] 17.006 ACC .707V 0.03% 17.007 MEME 17.008 MEMC V 0.50000% 1000pC #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 18.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 18.001 DISP TRANSDUCER SENSITIVITY for 1.00E+2 pc/MU. 18.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 18.002 DISP SCALE 1 MU/Volt. 18.003 DISP Set the UUT OPERATE. 18.004 5700 70.7mV 400H S 2W 18.005 IEEE [@1281]RDG?[D2000][I] 18.006 ACC .707V 0.03% 18.007 MEME 18.008 MEMC V 0.50000% 100pC #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 19.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 19.001 DISP TRANSDUCER SENSITIVITY for 1.00E+1 pc/MU. 19.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 19.002 DISP SCALE 10 MU/Volt. 19.003 DISP Set the UUT OPERATE. 19.004 5700 70.7mV 400H S 2W 19.005 IEEE [@1281]RDG?[D2000][I] 19.006 ACC .707V 0.03% 19.007 MEME 19.008 MEMC V 0.50000% 10pC #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 20.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 20.001 DISP TRANSDUCER SENSITIVITY for 1.00E+0 pc/MU. 20.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 20.002 DISP SCALE 100 MU/Volt. 20.003 DISP Set the UUT OPERATE. 20.004 5700 70.7mV 400H S 2W 20.005 IEEE [@1281]RDG?[D2000][I] 20.006 ACC .707V 0.03% 20.007 MEME 20.008 MEMC V 0.50000% 1pC #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 21.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 21.001 DISP TRANSDUCER SENSITIVITY for 1.00E-1 pc/MU. 21.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 21.002 DISP SCALE 1000 MU/Volt. 21.003 DISP Set the UUT OPERATE. 21.004 5700 70.7mV 400H S 2W 21.005 IEEE [@1281]RDG?[D2000][I] 21.006 ACC .707V 0.03% 21.007 MEME 21.008 MEMC V 0.50000% .1pC #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 22.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 22.001 DISP TRANSDUCER SENSITIVITY for 1.00E-2 pc/MU. 22.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 22.002 DISP SCALE 10000 MU/Volt. 22.003 DISP Set the UUT OPERATE. 22.004 5700 70.7mV 400H S 2W 22.005 IEEE [@1281]RDG?[D2000][I] 22.006 ACC .707V 0.03% 22.007 MEME 22.008 MEMC V 0.50000% .01PC #! Test Tol 0.003535, Sys Tol 0.0002121, TUR 16.667 (>= 4.00). 23.001 DISP Set the UUT to RESET. 23.002 HEAD {* Filter Cutoff Frequency Test *} 23.003 JMP 24.001 23.004 EVAL 24.001 DISP Using the UUT SELECT and arrow buttons, set the UUT 24.001 DISP TRANSDUCER SENSITIVITY for 1.00E+0 pc/MU. 24.002 DISP Using the UUT SELECT and arrow buttons, set the UUT 24.002 DISP SCALE 1000 MU/Volt. 24.003 DISP Set the UUT OPERATE. 24.004 IEEE [@1281]DB ON 24.005 MESS Adjust the Stimulus output until the 1281 reads 0.000dB. 24.006 5700 1V 1kH N 2W 24.007 MESS Adjust the Stimulus output until the 1281 reads -3.00dB. 24.008 5700 180kH 10.0% V 2W #! T.U.R. not calculated because MOD1 value not specified. 25.001 HEAD {* Frequency Response *} 25.002 JMP 26.001 25.003 EVAL 26.001 IEEE [@1281]DB OFF 26.002 5700 1V 400H S 2W 26.003 IEEE [@1281]RDG?[D2000][I] 26.004 MEME >1 26.005 IEEE [@1281]ACV AUTO,FILT10HZ 26.006 5700 1V 10H S 2W 26.007 IEEE [@1281]RDG?[D5000][I] 26.008 MEME 26.009 MEME <1 26.010 MEMC V 5.00% 10H #! Test Tol 0.05, Sys Tol 0.00065, TUR 76.923 (>= 4.00). 27.001 IEEE [@1281]ACV AUTO,FILT100HZ 27.002 5700 1V 1kH S 2W 27.003 IEEE [@1281]RDG?[D5000][I] 27.004 MEME 27.005 MEME <1 27.006 MEMC V 5.00% 1kH #! Test Tol 0.05, Sys Tol 8.2e-005, TUR 609.756 (>= 4.00). 28.001 5700 1V 10kH S 2W 28.002 IEEE [@1281]RDG?[D5000][I] 28.003 MEME 28.004 MEME <1 28.005 MEMC V 5.00% 10kH #! Test Tol 0.05, Sys Tol 8.2e-005, TUR 609.756 (>= 4.00). 29.001 5700 1V 50kH S 2W 29.002 IEEE [@1281]RDG?[D5000][I] 29.003 MEME 29.004 MEME <1 29.005 MEMC V 5.00% 50kH #! Test Tol 0.05, Sys Tol 0.00015, TUR 333.333 (>= 4.00). 30.001 5700 1V 100kH S 2W 30.002 IEEE [@1281]RDG?[D5000][I] 30.003 MEME 30.004 MEME <1 30.005 MEMC V 5.00% 100kH #! Test Tol 0.05, Sys Tol 0.00033, TUR 151.515 (>= 4.00). 31.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 1 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.